Guidance and manufacturer´s declaration –
Electromagnetic immunity for equipment and systems that
are not life-supporting
The Infrared Thermometer JPD-FR302 is intended for use
in the electromagnetic environment specified below. The
customer or the user of the Infrared Thermometer JPD-
FR302 should assure that it is used in such an environment.
Immunity
test
Electromagnetic
environment guidance
Radiated
RF
IEC
61000-4-3
3 V/m
80 MHz
to 2.5 GHz
Portable and mobile
RF communications
equipment should be used
no closer to any part of the
JPD-FR302, including
cables, than the
recommended separation
distance calculated from
the equation applicable to
the frequency of the
transmitter.
Recommended separation
distance
where p is the maximum
output power rating of the
transmitter in watts (W)
according to the transmitter
manufacturer and d is the
recommended separation
distance in metres (m).
Field strengths from fixed
RF transmitters, as
determined by an
Electromagnetic (a)
site survey, a should be
less than the compliance
level in each frequency
range (b). Interference may
occur in the vicinity of
equipment marked with
the following symbol:
IEC 60601
test level
3 V/m
Compliance
level
NOTE 1 At 80 MHz and 800 MHz, the higher frequency
range applies.
NOTE 2 These guidelines may not apply in all situations.
Electromagnetic is aected by absorption and reflection
from structures, objects and people.
a. Field strengths from fixed transmitters, such as base
stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast
and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment
due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field
strength in the location in which the JPD-FR302 is
used exceeds the applicable RF compliance level
above, the JPD-FR302 should be observed to verify
normal operation. If abnormal performance is observed,
additional measures may be necessary, such as re-
orienting or relocating the JPD-FR302.
b. Over the frequency range 150 kHz to 80 MHz, field
strengths should be less than 3 V/m.
Recommended separation distances between portable and
mobile RF communications equipment and the EQUIPMENT or
SYSTEM -for EQUIPMENT and SYSTEMS that are not
LIFE-SUPPORTING
The Infrared Thermometer JPD-FR302 is intended for use in an
electromagnetic environment in which radiated RF disturbances
are controlled. The customer or the user of the Infrared
Thermometer JPD-FR302 can help prevent electromagnetic
interference by maintaining a minimum distance between
portable and mobile RF communications equipment
(transmitters) and the Infrared Thermometer JPD-FR302 as
recommended below, according to the maximum output power
of the communications equipment.
Rated maximun
output power
of transmitter W
Separation distance according to frequency
of transmitter m
0.01
For transmitters rated at a maximum output power not listed
above, the recommended separation distance d in metres (m)
can be estimated using the equation applicable to the frequency
of the transmitter, where P is the maximum output power rating
of the transmitter in watts (W) according to the transmitter
manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the
higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations.
Electromagnetic propagation is aected by absorption and
reflection from structures, objects and people.
80 MHz to 800 MHz 800 MHz to 2.5 GHz
0.12 0.23
0.1 0.38 0.73
11.2 2.3
10 3.8 7.3
100 12 23
CAUTION.
IP22
BF type applied part.
Classification for water ingress and particulate matter.
Indicates this device is subject to the Waste Electrical and
Electronic Equipment Directive in the European Union. To
protect the environment dispose of useless device at appropriate
collection sites according to national or local regulations.
Please read instructions. Battery recycling.
Maufactured by: Shenzhen Jumper Medical Equipment Co., Ltd.
Address: D Building, No. 71, Xintian Road, Fuyong Street,
Baoan, Shenzhen, Guangdong, China
Distributed by: Mayborn (UK) Limited, Balliol Business Park,
Newcastle upon Tyne, NE12 8EW.
Mayborn France Sarl, 92100, France.
Mayborn USA Inc. CT 06901, USA.
MedPath GmbH
Mies-van-der-Rohe-Strasse 8, 80807 Munich, Germany.
Information on the Disposal of Waste Electrical and Electronic
Equipment (WEEE)
This symbol means that your product shall be disposed of
separately from household waste in accordance with local laws
and regulations. When this product is ready for disposal, take it
to a collection point designated by local authorities. Separating
and recycling of your product at the time of disposal will
help conserve natural resources and ensure that the product
is recycled in a manner that protects human health and the
environment. Penalties may be applicable for the incorrect
disposal of this product in accordance with national legislation.