BD ClipVac 5506A Instructions For Use Manual

Tip
Instructions For Use Manual
5500E / 5506A / 5506E / 5506U / 5506K
Instructions For Use
ClipVac™ Presurgical Hair Removal System
Español . . . . . . .54
Português . . . . .67
Nederlands. . . .80
Svenska . . . . . . .93
es
pt
nl
sv
Deutsch . . . . . . .28
Italiano . . . . . . .41
Français. . . . . . .15
fr
de
it
English . . . . . . . . . 2
en
Türkçe. . . . . . . 158
Polski. . . . . . . . 171
Česky . . . . . . . 184
tr
pl
cz
Dansk . . . . . . . 106
da
Greek . . . . . . . 145
el
Suomi . . . . . . . 119
Norsk . . . . . . . 132
fi
no
2
Intended Use
The intended use of the BD ClipVac™ system is to vacuum clipped hair and airborne contaminants generated from the clipping process. It should be used
with the BD Surgical Clipper (REF 5513E). Only trained healthcare personnel should use BD ClipVac™ system and the BD Surgical Clippers. If the
BD Surgical Clippers and BD ClipVac™ system are being used for the first time, see the training materials provided by BD.
How Supplied:
The BD ClipVac™ Vacuum Unit (REF 5500E) is a non-sterile reusable device. Included in this box:
1. One vacuum unit
2. IFU
The BD ClipVac™ Charging Adapter & Battery will be included in a separate box (REF 5506A/5506E/5506U/5506K) that contains:
1. One battery
2. One charging adapter (charger, power supply & power cord)
Note: The BD ClipVac™ Disposable Filter Assembly is packaged and sold separately (REF 5575).
Limitations on Reprocessing:
Repeated cleaning has minimal effect on the system. End of life is determined by wear and inadvertent damage due to use.
BD ClipVac™ Parts Identification:
REF 5500E: Vacuum Unit
REF 5505: Replacement Battery
REF 5506A/5506E/5506U/5506K: Charging Adapter & Battery
REF 5575: Disposable Filter Assembly
en
3
Step 1. Remove the battery from charging
stand and review the LCD display to
ensure battery has sufficient charge for
the procedure prior to use. See Charging
the Battery section.
Step 2. To insert battery in vacuum unit,
retract battery latch and place in open
slot with the battery tab opposite the
battery latch.
Note: Battery will be properly seated
when latch mechanism is slightly over
the battery.
Step 3. Remove the disposable filter
assembly from package. Uncoil tubing
by gripping near the nozzle and filter
body , gently pulling taut to remove any
twist and straighten tubing.
Step 5. Place nozzle of disposable filter
assembly over the clipper after the
surgical blade has been installed on the
clipper handle. Nozzle will click into place
when properly seated to the clipper.
Ensure the retention knob on either
side of the nozzle is properly seated prior
to use.
Step 4. Place the filter body onto the
vacuum port end of the vacuum unit.
Note: The filter body will rotate freely
when attached.
1
2
3
5
4
6
4
Step 7. While holding loose skin taut, clip
hair using a slow steady forward motion
avoiding downward pressure.
Note: The bottom side of the clipper
blade should remain flat and gently rest
on the surface of the skin while clipping.
To ensure hair is properly collected by the
vacuum, never tilt the blade edge into
the skin.
Note: For optimal performance, use on
dry hair.
Step 8. When clipping is complete, turn
OFF unit and gently twist the filter body
to remove disposable filter assembly
away from the vacuum unit.
Step 9. Discard filter body into general
waste and recycle nozzle and tubing
according to local regulations. To detach
the filter body from the tubing, pull with
a gentle twist.
Step 10. After disinfecting and/or manual
cleaning, place vacuum unit in the
designated area for next use. If necessary,
return battery to the charging adapter.
Step 6. Turn ON the vacuum unit by
flipping the switch .
Note: Ensure BD Surgical Clipper is turned
ON prior to starting clipping procedure.
SN: 00 00 000
Note: The BD ClipVac™ Vacuum Unit may be placed on the bed next to the patient or
secured to the bedrail.
7
Charging notes:
Maximum time required for a charging adapter to charge an empty battery shall be 4 hours.
Charge at room temperature of +41 °F (+5 °C) to +95 °F (+35 °C)
Avoid charging in direct sunlight or near radiating heat.
If not used for extended periods of time, charge battery every six (6) months to preserve battery life.
The lithium-ion battery must be disposed of in accordance with local regulations or by using the RBRC 1-800 number listed on the battery to locate
your local recycling point. Never incinerate the lithium battery. Do not dispose of the battery with general waste.
Disinfection Instructions:
The BD ClipVac™ Vacuum Unit should be thoroughly disinfected, after each use, using one of the recommended disinfectant solutions listed below:
Isopropyl Alcohol/Quaternary Ammonia solution (e.g., validated disinfectant PDI Super Sani-Cloth
®
) containing the active ingredients:
º
0.25% n-Alkyl (68% C
12
, 32% C
14
) dimethyl ethylbenzyl ammonium chloride
º
0.25% n-Alkyl (60% C
14
, 30% C
16
, 5% C
12
, 5% C
18
) dimethyl benzyl ammonium chloride
º
55.00% Isopropyl Alcohol
Chlorinated bleach solution (e.g. validated disinfectant PDI Sani-Cloth
®
Bleach) containing the active ingredient:
º
0.63% sodium hypochlorite (i.e., 6300 ppm sodium hypochlorite)
1. Thoroughly apply to all surfaces of the BD ClipVac™ Vacuum Unit and maintain a wet surface for at least:
a. 3 minutes for Isopropyl Alcohol/Quaternary Ammonia solutions or
b. 4 minutes for Chlorinated bleach solution.
2. Surfaces must remain visibly wet with the disinfectant for the complete timeframe indicated above.
3. Repeat the disinfection steps above if the unit dries prior to the allotted timeframe.
4. Allow the BD ClipVac™ Vacuum Unit to air dry completely prior to storage.
5
Charging the Battery
Plug the charging cord into the charging adapter. A quick red and green LED flashing is
an indication that charger is ON. Place the battery into the battery bay ensuring that the
5-way connector is fully seated. The LCD fuel gauge on the battery will begin flashing to
indicate it is receiving a charge. The LED indicators on the front of the charging adapter
will provide the charging status. Below are the LED indicators on the charging adapter:
Green Flashing………….Battery is charging
Green Solid………………..Battery is fully charged
Red Solid…………………….Battery failure, replace battery
Fuel Gauge LCD display on top of the battery:
5 Bars = 80% to 100% charged
4 Bars = 60% to 79% charged
3 Bars = 40% to 59% charged
2 Bars = 20% to 39% charged
1 Bar = 0% to 19% charged
6
If the device comes in contact with bodily fluid, the following manual cleaning instructions can be used:
1. Using enzymatic/neutral pH detergent, prepare per the manufacturer’s instructions utilizing tap water entailing a temperature range of 27°C to 44°C
(81°F to 111°F).
2. Do not submerge or spray the device with cleaning solution.
3. Use a soft, clean cloth saturated with the cleaning solution (not dripping) to manually clean all external contaminated surfaces of the device for a
minimum of 5 minutes. Be careful not to cause any liquid to drip into the charger or vacuum. In conjunction with this step, use an appropriately sized
brush (standard soft bristle toothbrush and lumen brush) to remove all visible soil from the device; paying particular attention to cracks, crevices and
other hard to clean areas. Note: It is recommended that enzymatic or neutral pH detergent solution should be changed when it becomes grossly
contaminated (bloody and/or turbid).
4. Rinse thoroughly by using a soft, clean cloth saturated with lukewarm (27°C - 44°C) tap water (not dripping) for a minimum of 30 seconds. Repeat 2
additional times for a total of 3 total rinse cycles. Be careful not to cause any liquid to drip into the charger or vacuum.
5. Dry the device with a clean, lint-free towel.
6. Visually examine the device for cleanliness.
7. If visible soil remains, repeat cleaning procedure until the device is thoroughly clean.
Inspection/Maintenance
The BD ClipVac™ Vacuum Unit requires no routine maintenance aside from cleaning system after each patient use. If a replacement is needed,
contact your local BD representative or customer service. Service or attempted repair performed by unqualified personnel may result in a risk of injury,
electric shock, fire, or permanent damage to the equipment and will void any warranty.
Never use BD ClipVac™ Vacuum Unit or a battery that appears to be defective, damaged, or not working properly.
Limited Warranty:
The BD ClipVac™ Presurgical Hair Removal System from BD (REFs 5500E, 5505, 5506A/5506E/5506U/5506K) is warranted against defects in material
and workmanship under normal use for a period of two (2) years from the date of unit placement. If this product is determined to be defective with
the warranty period, it will be repaired or replaced without charge by BD, if you contact your local representative or customer service representative.
They will provide the proper return authorization. Please include a description of the defect or damage with the returned unit.
This warranty does not cover the Disposable Filter Assembly (REF 5575) nor does it cover damage caused by misuse, negligence, accident, or abuse,
alteration or modification of the unit by anyone other than BD or its authorized agent, or affixation of any accessory not manufactured or authorized
by BD, or used in combination with other surgical clippers.
BD is not liable in either tort or contract for any cost, expense, loss, or damage whether direct, consequential, special, collateral, or incidental arising
out of the use, misuse, or inability to use this product. Except as expressly provided above, BD disclaims all warranties including, but not limited to the
implied warranties of merchantability and fitness for particular purpose.
7
Disposal:
Disposable assembly can be discarded into general waste or per hospital protocol.
The nozzle and tubing may be recycled. To remove the filter body from the tubing, simply pull the tube from the filter body.
The lithium-ion battery must be disposed of in accordance with local regulations or by using the RBRC 1-800 number listed on the battery to locate
your local recycling point. Never incinerate the lithium battery. Do not dispose of the battery with general waste.
Caution:
Use the BD ClipVac™ system as described in the Operating Instruction. Avoid direct patient contact with the nozzle. Direct patient contact should be
with the bottom side of the clipper blade.
Never use BD ClipVac™ Vacuum Unit, battery or charging adaptor that appears to be defective, damaged, or not working properly.
Do not cover vacuum unit to prevent obstructing air flow as this may cause overheating of vacuum unit.
Do not re-use disposable filter assembly, it is intended to be single use only.
Clean the vacuum unit after each patient use, refer to cleaning and disinfecting instructions.
Do not submerse or spray the vacuum unit, battery, or charging adapter with any liquid.
Do not sterilize the vacuum unit, battery or charging adapter.
Do not short the battery terminals.
Avoid excessive physical shock or vibration to the battery.
Do not place the battery and charging adapter in a location where it is subjected to direct sunlight and keep away from other external heat sources.
Do not dispose the battery with general waste. Never incinerate the lithium ion battery.
Do not replace charging adapter cord. Use only the battery and charging adapter supplied by BD.
Do not attempt to modify or repair the vacuum unit, battery, or charging adapter as this can result in damage of the BD ClipVac™ system and will void
the warranty.
Symbol Legend:
Switch “Off”
(Power)
Switch “On
(Power)
Do not reuse
Non-ionizing
electromagnetic radiation
Temperature limit
-20°C
+50°C
Atmospheric
pressure limitation
50 kPa
106 kPa
Humidity limitation
0%
80%
Information for Users on Collection and Disposal of Old Equipment and used Batteries
These symbols on the products, packaging, and/or accompanying documents mean that used electrical and electronic products and batteries
should not be mixed with general household waste. For proper treatment, recovery and recycling of old products and used batteries, please
take them to applicable collection points, in accordance with your national legislation and the Directives 2012/19/EU and 2006/66/EC. By
disposing of these products and batteries correctly, you will help to save valuable resources and prevent any potential negative effects on
human health and the environment which could otherwise arise from inappropriate waste handling.
For more information about collection and recycling of old products and batteries, please contact your local municipality, your waste disposal
service or the point of sale where you purchased the items. Penalties may be applicable for incorrect disposal of this waste, in accordance with
national legislation.
Information on Disposal in other Countries outside the European Union
These symbols are only valid in the European Union. If you wish to discard these items, please contact your local authorities or dealer and ask for the
correct method of disposal.
Catalog Number
Batch Code
Manufacturer
Date of Manufacture
8
Type BF applied part
Consult instructions
for use
Direct current
Authorized representative in
the European Community
9
BD ClipVac™ Vacuum Unit Mode
of Operation
Continuous
BD ClipVac™ Disposable Filter
Assembly Mode of Operation
Single Use
BD ClipVac™ Vacuum Unit 14.4VDC - 2.8A
BD ClipVac™ Vacuum Unit Weight 725g/25.6oz
Battery Type Lithium ion with LCD Display
Battery Weight 234g/8.25 oz.
Full Charge Time Approximately 3.5 hours
Continuous Run Time
Approximately 75 minutes
on a full charge
Battery Life Minimum 260 charges
Power Supply 100 – 240 VAC, 24V 2.71A DC
Ambient Operating
Temperature, Humidity, and
Atmospheric Pressure
10°C to 40°C (50°F to 104°F)
15% to 95% RH
non-condensing
70 - 106kpa.
Storage Temperature, Humidity,
and Atmospheric Pressure
-20°C to 50°C ≤ 80% RH
-4°F to 122°F ≤ 80% RH
50 - 106kpa
BD ClipVac™ Vacuum Unit Specifications:
Weight 235g/8 oz.
Height 58mm/2-1/4”
Length 180mm/7”
Width 92mm/3-1/2"
Mating Connector
5-blade standard battery
connector
Communications Compliances
System Management Bus
Rev. 1.0, Smart Battery Data
Specification rev 1.0 and Smart
Battery Charger specification
Rev. 1.0 compliant
Power Supply
100-240VAC, 24V, 2.71A DC,
119x38x60mm, 500g/18oz,
UL Listed, CE Compliant, CEC
Compliant
Mains Cord
5506A - Type B
5506E - Type F
5506K - Type I
5506U - Type G
Charging Adapter Specification:
EMC Information of REF 5500
REF 5500E needs special precautions regarding EMC and needs to be used according to the following EMC information. REF 5500E complies with the
EMC requirements of the IEC 60601-1-2.
The BD ClipVac™ system is intended for use in the electromagnetic environment specified in the following tables.
Warning:
REF 5500E should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is necessary, REF 5500E should be observed to
verify normal operation in the configuration in which it will be used.
10
REF 5500E has been tested and found to comply with the limits for the medical devices to the IEC 6060112. These limits are designed to provide
reasonable protection against harmful interference. This equipment generates uses and can radiate radio frequency energy and, if not installed
and used in accordance with the instructions, may cause harmful interference to other devices in the vicinity. However, there is no guarantee that
interference will not occur in a particular installation. If this equipment does cause harmful interference to the other devices which can be determined
by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures;
Reorient or relocate the receiving device.
Increase the separation between the equipment.
Connect the equipment into an outlet on a circuit different from that to which the other device(s) are connected.
Consult the manufacturer or field service for help.
11
Guidance and manufacturer’s declaration — electromagnetic emissions
REF 5500E is intended for use in the electromagnetic environment specified below.
The customer or the user of REF 5500E/5506 series should insure that it is used in such an environment.
Emission test Compliance Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
REF 5500E uses RF energy only for its internal function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
REF 5500E is suitable use in all establishments, including domestic establishments and those
directly connected to the public low-voltage power supply network that supplies buildings used
for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
12
Guidance and manufacturer’s declaration — electromagnetic immunity
REF 5500E/5506 series is intended for use in the electromagnetic environment specified below.
The customer or the user of REF 5500E/5506 series should insure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment - guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
±8 kV contact
±15 kV air
±8 kV contact
±15
1
kV air
±8
2
kV air
Floors should be wood, concrete or ceramic tile. If floors are covered
with synthetic material, the relative humidity should be at least
30%.
Electrical fast transient/burst
IEC 61000-4-4
±2 kV for power supply lines
±1 kV for input/output lines
±2 kV for power supply lines
±1 kV for input/output lines
Mains power quality should be that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
±1 kV for differential mode
±2 kV common mode
±1 kV for differential mode
±2 kV common mode
Mains power quality should be that of a typical commercial or
hospital environment.
Voltage dips, short interruptions
and voltage variations on power
supply input lines
IEC 61000-4-11
<0% (>100% dip in )
for 0.5 cycle (phase angle
0°,45°,90°,135°,180°,225°,270°)
0% (100% dip in )
for 1 cycles
70% (30% dip in )
for 25/30 cycles
<0% (>100% dip in ) for 250
cycle
<0% (>100% dip in )
for 0.5 cycle (phase angle
0°,45°,90°,135°,180°,225°,270°)
0% (100% dip in )
for 5 cycles
70% (30% dip in )
for 25 cycles
<0% (>100% dip in ) for 250
cycle
Mains power quality should be that of a typical commercial or
hospital environment.
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m 30 A/m
Power frequency magnetic fields should be at levels characteristic
of a typical location in a typical commercial or hospital
environment.
REF 5500E - Vacuum unit and charger compliance level to ±15 kV air discharge.
REF 5506A/5506E/5506U/5506K - Battery compliance level to ±8 kV air discharge. Battery fuel gauge LCD display may be damaged at ±15 kV air. But this is not a hazard to the patient.
Note: is the a.c. mains voltage prior to application of the test level
13
Guidance and manufacturer’s declaration — electromagnetic immunity
REF 5500E/5506 series is intended for use in the electromagnetic environment specified below.
The customer or the user of REF 5500E/5506 series should insure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment - guidance
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
6 Vrms
150 kHz to 80 MHz (ISM and
amateur radio bands)
3 Vrms (150 kHz to 80 MHz)
6 Vrms (150 kHz to 80 MHz)
Portable and mobile RF communications equipment should be used no closer
to any part of REF 5500E, including cables, than the recommended separation
distance calculated from the equation applicable to the frequency of the
transmitter.
Recommended separation distance
d = 1.2 √P
d = 1.2 √P 80 MHz to 800 MHz
d = 2.3 √P 800 MHz to 2.7 GHz
Where P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer and d is the recommended separation
distance in meters (m).
Field strength from fixed RF transmitters, as determined by an electromagnetic
site survey, should be less than the compliance level in each frequency range.
Interference may occur in the vicinity of equipment marked with the following
symbol:
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.7 GHz
3 V/m
80 MHz to 2.7 GHz
Enclosure port immunity IEC
61000-4-3
27 V/m, 385MHz
28 V/m, 450MHz
9 V/m, 710 MHz to 780 MHz
28 V/m, 810 MHz to 2450 MHz
9 V/m, 5240 MHz to 5785 MHz
27 V/m, 385MHz
28 V/m, 450MHz
9 V/m, 710 MHz to 780 MHz
28 V/m, 810 MHz to 2450 MHz
9 V/m, 5240 MHz to 5785 MHz
d = 6/E√P
380MHz to 5.8GHz
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the
measured field strength in the location in which REF 5500E is used exceeds the applicable RF compliance level above, REF 5500E should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as reorienting or relocating REF 5500E.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
14
Recommended separation distance between portable and mobile RF communications equipment and REF 5500E.
REF 5500E/5506 series is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of
REF 5500E/5506 series can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications
equipment (transmitters) and REF 5500E/5506 series as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output power of
transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
d = 1.2 √P
80 MHz to 800 MHz
d = 1.2 √P
800MHz to 2.7GHz
d = 2.3 √P
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation
applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects
and people.
Utilisation prévue
L'utilisation prévue du système BD ClipVac™ est d'aspirer les poils et cheveux tondus et les contaminants en suspension dans l'air résultant de la tonte.
Le système doit être utilisé avec la tondeuse chirurgicale BD (RÉF.5513E). Seul le personnel de santé qualifié peut utiliser le système BD ClipVac™ et les
tondeuses chirurgicales BD. Lors de la première utilisation des tondeuses chirurgicales BD et du système BD ClipVac™, se reporter aux documents de
formation fournis par BD.
Présentation :
L'unité d'aspiration BD ClipVac™ (RÉF.5500E) est un appareil réutilisable non stérile. Contenu de la boîte:
1. Une unité d'aspiration
2. Instructions d'utilisation
L'adaptateur de charge et la batterie du système ClipVac™ BD sont inclus dans une boîte séparée (RÉF.5506A/5506E/5506U/5506K) qui contient:
1. Une batterie
2. Un adaptateur de charge (chargeur, bloc d'alimentation et cordon d'alimentation électrique)
Remarque: le filtre à usage unique du système BD ClipVac™ est emballé et vendu séparément (RÉF. 5575).
Limites du retraitement:
Le nettoyage répété a un impact limité sur le système. La durée de vie des instruments est déterminée par l'usure et les dommages accidentels résultant
de leur utilisation.
Identification des pièces du système BD ClipVac™:
RÉF. 5500E: Unité d'aspiration
RÉF. 5505: Batterie de rechange
RÉF. 5506A/5506E/5506U/5506K : Adaptateur de charge et batterie
RÉF. 5575: Filtre à usage unique
fr
15
1
2
3
5
4
6
Étape1. Retirer la batterie du support de
charge et consulter l'écran LCD pour
s'assurer que la batterie est suffisamment
chargée pour la procédure avant d'utiliser
l'instrument. Voir la section Chargement
de la batterie.
Étape2. Pour insérer la batterie dans
l'unité d'aspiration, rétracter le loquet de
la batterie et placer la batterie dans
le logement ouvert, en s'assurant que
la languette de batterie est opposée au
loquet de la batterie.
Remarque: la batterie est correctement
insérée lorsque le mécanisme de
verrouillage se trouve légèrement au-
dessus de la batterie.
Étape3. Retirer le filtre à usage unique
de son emballage. Dérouler le tube en
le tenant près de l'embout et du corps
de filtre , puis tirer légèrement dessus
pour éliminer les éventuels tortillons et le
redresser.
Étape5. Placer l'embout du filtre à usage
unique sur la tondeuse après avoir installé
la lame chirurgicale sur la poignée de
la tondeuse. L'embout s'enclenche en
place dès qu'il est correctement installé
sur la tondeuse. S'assurer que le bouton
de rétention qui se trouve de chaque
côté de l'embout est correctement installé
avant d'utiliser l'instrument.
Étape4. Placer le corps du filtre sur le port
d'aspiration de l'unité d'aspiration.
Remarque: le corps du filtre doit pouvoir
tourner librement lorsqu'il est fixé.
16
Step 8. When clipping is complete, turn
OFF unit and gently twist the filter body
to remove disposable filter assembly
away from the vacuum unit.
SN: 00 00 000
Remarque : l'unité d'aspiration du système BD ClipVac™ peut être placée sur le lit à côté du
patient ou fixée à l'armature du lit.
7
Étape7. Tout en tendant la peau, tondre les
poils en eectuant un mouvement lent et
continu vers l'avant et en évitant d'exercer une
pression vers le bas.
Remarque: la partie inférieure de la lame
de la tondeuse doit rester à plat et reposer
légèrement sur la surface de la peau pendant
la tonte. Pour s'assurer que les poils sont bien
aspirés, ne jamais incliner le bord de la lame
vers la peau.
Remarque: pour obtenir des performances
optimales, utiliser sur une peau et des poils secs.
Étape6. Allumer l'unité d'aspiration en
appuyant sur le bouton .
Remarque: s'assurer que la tondeuse
chirurgicale BD est bien allumée avant de
procéder à la tonte.
Étape9. Éliminer le corps du filtre
comme un déchet classique et recycler
l'embout et le tube conformément à la
réglementation locale. Pour détacher le
corps du filtre du tube, tirer dessus en le
tournant légèrement.
Étape10. Après désinfection et/
ou nettoyage manuel, placer l'unité
d'aspiration dans la zone désignée jusqu'à
la prochaine utilisation. Si nécessaire,
remettre la batterie dans l'adaptateur
de charge.
17
Remarques sur le chargement:
La durée maximale nécessaire à un adaptateur de charge pour charger une batterie vide est de 4heures.
Chargement à température ambiante (entre +5°C et +35°C).
Éviter de charger la batterie à la lumière directe du soleil ou à proximité d'une chaleur rayonnante.
En cas d'inutilisation pendant une période prolongée, charger la batterie tous les six (6) mois pour préserver l'autonomie de la batterie.
La batterie au lithium-ion doit être mise au rebut conformément à la réglementation locale ou dans le point de recyclage identifié à l'aide du numéro RBRC 1-800
qui figure sur la batterie. Ne jamais incinérer la batterie au lithium. Ne pas éliminer la batterie avec les déchets classiques.
Instructions de désinfection:
L'unité d'aspiration du système BD ClipVac™ doit être soigneusement désinfectée, après chaque utilisation, en utilisant l'une des solutions désinfectantes
recommandées ci-dessous:
Solution d'alcool isopropylique/ammonium quaternaire (ex.: désinfectant autorisé PDI Super Sani-Cloth
®
) contenant les ingrédients actifs suivants:
º
0,25% de chlorure de n-alkyl (68% de C
12
, 32% de C
14
) diméthyléthylbenzyl ammonium
º 0,25% de chlorure de n-alkyl (60% de C
14
, 30% de C
16
, 5% de C
12
, 5% C
18
) diméthylbenzyl ammonium
º 55% d'alcool isopropylique
Solution d'eau de javel chlorée (ex.: désinfectant autorisé PDI Sani-Cloth
®
Bleach) contenant l'ingrédient actif suivant:
º
0,63% d'hypochlorite de sodium (c.-à-d. 6300ppm d'hypochlorite de sodium)
1. Appliquer soigneusement sur toutes les surfaces de l'unité d'aspiration du système BD ClipVac™ et conserver une surface humide pendant au moins:
a. 3minutes pour les solutions d'alcool isopropylique/ammonium quaternaire ou
b. 4minutes pour la solution d'eau de javel chlorée.
2. Les surfaces doivent rester visiblement humidifiées par le désinfectant pendant l'intégralité des durées indiquées ci-dessus.
3. Répéter les étapes de désinfection ci-dessus si l'unité sèche avant le délai imparti.
4. Laisser l'unité d'aspiration du système BD ClipVac™ sécher entièrement à l'air libre avant de la ranger.
Chargement de la batterie
Brancher le cordon de charge sur l'adaptateur de charge. Lorsque le voyant rouge et vert
clignote rapidement, cela signifie que le chargeur est allumé. Placer la batterie dans son
compartiment, en veillant à ce que le connecteur à 5 broches soit bien en place. La jauge
LCD située sur la batterie se met à clignoter pour indiquer que la batterie est en cours de
chargement. Les voyants à LED situés sur la façade de l'adaptateur de charge indiquent l'état
de charge. Les voyants à LED qui se trouvent sur l'adaptateur de charge sont les suivants:
Vert clignotant ......Batterie en cours de chargement
Vert fixe ............. Batterie pleine
Rouge fixe...........Défaillance de la batterie, procéder à son remplacement
Écran LCD de la jauge sur le dessus de la
batterie:
5barres = 80% à 100% de charge
4barres = 60% à 79% de charge
3barres = 40% à 59% de charge
2barres = 20% à 39% de charge
1barre = 0% à 19% de charge
18
Si l'appareil entre en contact avec des fluides corporels, les instructions de nettoyage manuel suivantes peuvent être suivies:
1. À l'aide d'un détergent enzymatique/à pH neutre, préparer une solution conformément aux instructions du fabricant, en utilisant de l'eau du robinet à une
température comprise entre 27°C et 44°C.
2. Ne pas immerger l'appareil dans la solution de nettoyage ou l'asperger avec celle-ci.
3. Utiliser un chion doux et propre, imprégné de solution de nettoyage (mais qui ne goutte pas), pour nettoyer manuellement toutes les surfaces extérieures
contaminées de l'appareil pendant au moins 5minutes. Veiller à ce qu'aucune goutte de liquide ne pénètre dans le chargeur ou l'unité d'aspiration.
Parallèlement, utiliser une brosse de taille adaptée (brosse à dents classique à poils souples et écouvillon) pour éliminer toute salissure visible de l'appareil, en
faisant particulièrement attention aux fissures, aux fentes et aux autres zones diciles d'accès. Remarque: il est conseillé de remplacer la solution détergente
enzymatique ou à pH neutre lorsqu'elle est excessivement souillée (sanglante et/ou trouble).
4. Rincer soigneusement à l'aide d'un chion doux et propre, imprégné d'eau tiède (27°C - 44°C), pendant au moins 30secondes. Répéter l'opération 2fois, pour un
total de 3cycles de rinçage complets. Veiller à ce qu'aucune goutte de liquide ne pénètre dans le chargeur ou l'unité d'aspiration.
5. Sécher le dispositif à l'aide d'une serviette propre et qui ne peluche pas.
6. Examiner visuellement le dispositif pour vérifier sa propreté.
7. Si des souillures sont encore présentes, répéter la procédure de nettoyage jusqu'à ce que le dispositif soit parfaitement propre.
Inspection/entretien
L'unité d'aspiration du BD ClipVac™ ne nécessite aucun entretien de routine, en dehors du système de nettoyage, après chaque utilisation sur un patient. Si un
remplacement est nécessaire, contacter votre représentant BD local ou le service clientèle. Tout entretien ou toute tentative d'entretien par du personnel non
qualifié peut présenter un risque de blessures, d'électrocution, d' incendie ou de dommages irréversibles sur l'appareil, et entraîner l'annulation de la garantie.
Ne jamais utiliser l'unité d'aspiration du système BD ClipVac™ ou une batterie qui semble défectueuse, endommagée ou qui ne fonctionne pas correctement.
Garantie limitée:
Le système de rasage préchirurgical BD ClipVac™ de BD (RÉF.5500E, 5505, 5506A/5506E/5506U/5506K) est garanti contre les défauts matériels et les vices
de fabrication, dans des conditions normales d'utilisation, pendant une période de deux (2) ans à compter de la date de mise en place de l'unité. Si ce produit
est jugé défectueux pendant la période de garantie, il sera alors réparé ou remplacé gratuitement par BD après avoir contacté votre représentant local ou un
représentant du service clientèle. Celui-ci vous transmettra l'autorisation de retour nécessaire. Une description du défaut ou du dommage subi par l'unité renvoyée
doit être fournie.
Cette garantie ne couvre pas le filtre à usage unique (RÉF.5575) ou les dommages provoqués par un abus, une négligence, un accident ou une modification de
l'appareil par une personne autre que BD ou un agent agréé, ou par l'ajout d'un quelconque accessoire qui ne serait pas fabriqué ou autorisé par BD ou qui serait
utilisé en combinaison avec d'autres tondeuses chirurgicales.
BD décline toute responsabilité, délictuelle ou contractuelle, en cas de coûts, de dépenses, de pertes ou de dommages directs, indirects, spéciaux, collatéraux
ou accessoires engendrés par l'utilisation, le mauvais usage, ou l'incapacité à utiliser ce produit. Sauf disposition expresse ci-dessus, BD décline toute garantie, y
compris, sans toutefois s'y limiter, toute garantie implicite de qualité marchande et d' adéquation à un usage particulier.
19
Mise au rebut:
Les dispositifs à usage unique peuvent être éliminés avec les déchets classiques ou conformément au protocole de l'hôpital.
L'embout et les tubes peuvent être recyclés. Pour retirer le corps du filtre du tube, tirer simplement sur le tube pour le séparer du corps du filtre.
La batterie au lithium-ion doit être mise au rebut conformément à la réglementation locale ou dans le point de recyclage identifié à l'aide du numéro
RBRC 1-800 qui figure sur la batterie. Ne jamais incinérer la batterie au lithium. Ne pas éliminer la batterie avec les déchets classiques.
Attention:
Utiliser le système BD ClipVac™ conformément aux instructions d'utilisation. Éviter tout contact direct entre le patient et l'embout. Le contact direct
avec le patient doit s'eectuer avec la partie inférieure de la lame de la tondeuse.
Ne jamais utiliser l'unité d'aspiration du système BD ClipVac™, la batterie ou un adaptateur de charge qui semble défectueux(se), endommagé(e) ou
qui ne fonctionne pas correctement.
Ne jamais couvrir l'unité d'aspiration; cela pourrait empêcher la circulation de l'air et provoquer une surchaue de l'unité d'aspiration.
Ne pas réutiliser le filtre à usage unique, il est destiné à n'être utilisé qu'une seule fois.
Nettoyer l'unité d'aspiration après chaque utilisation sur un patient en respectant les instructions de nettoyage et de désinfection.
Ne jamais immerger l'unité d'aspiration, la batterie ou l'adaptateur de charge dans un liquide ou l'asperger avec celui-ci.
Ne pas stériliser l'unité d'aspiration, la batterie ou l'adaptateur de charge.
Ne pas court-circuiter les bornes de la batterie.
Éviter tout choc physique ou toute vibration excessive de la batterie.
Ne pas placer la batterie et l'adaptateur de charge dans un endroit exposé aux rayons directs du soleil et les éloigner de toute autre source de chaleur
externe.
Ne pas éliminer la batterie avec les déchets classiques. Ne jamais incinérer la batterie au lithium-ion.
Ne pas remplacer le cordon de l'adaptateur de charge. Utiliser uniquement la batterie et l'adaptateur de charge fournis par BD.
Ne pas essayer de modifier ou de réparer l'unité d'aspiration, la batterie ou l'adaptateur de charge; cela risquerait d'endommager le système
BD ClipVac™ et d'annuler la garantie.
20
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120
  • Page 121 121
  • Page 122 122
  • Page 123 123
  • Page 124 124
  • Page 125 125
  • Page 126 126
  • Page 127 127
  • Page 128 128
  • Page 129 129
  • Page 130 130
  • Page 131 131
  • Page 132 132
  • Page 133 133
  • Page 134 134
  • Page 135 135
  • Page 136 136
  • Page 137 137
  • Page 138 138
  • Page 139 139
  • Page 140 140
  • Page 141 141
  • Page 142 142
  • Page 143 143
  • Page 144 144
  • Page 145 145
  • Page 146 146
  • Page 147 147
  • Page 148 148
  • Page 149 149
  • Page 150 150
  • Page 151 151
  • Page 152 152
  • Page 153 153
  • Page 154 154
  • Page 155 155
  • Page 156 156
  • Page 157 157
  • Page 158 158
  • Page 159 159
  • Page 160 160
  • Page 161 161
  • Page 162 162
  • Page 163 163
  • Page 164 164
  • Page 165 165
  • Page 166 166
  • Page 167 167
  • Page 168 168
  • Page 169 169
  • Page 170 170
  • Page 171 171
  • Page 172 172
  • Page 173 173
  • Page 174 174
  • Page 175 175
  • Page 176 176
  • Page 177 177
  • Page 178 178
  • Page 179 179
  • Page 180 180
  • Page 181 181
  • Page 182 182
  • Page 183 183
  • Page 184 184
  • Page 185 185
  • Page 186 186
  • Page 187 187
  • Page 188 188
  • Page 189 189
  • Page 190 190
  • Page 191 191
  • Page 192 192
  • Page 193 193
  • Page 194 194
  • Page 195 195
  • Page 196 196
  • Page 197 197
  • Page 198 198

BD ClipVac 5506A Instructions For Use Manual

Tip
Instructions For Use Manual